Vulnerability in Hospira Plum A+3 Infusion System

CVE-2015-3953

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 9.8 (Critical). Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.

EPSS exploit prediction

EPSS 0.020, 80.8th percentile (scored ); unchanged over 30 days. About EPSS.

Affected products

Weakness classification (CWE)

References