Information disclosure in Hospira Plum A+3 Infusion System

CVE-2015-3952

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 7.5 (High). Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.

EPSS exploit prediction

EPSS: 0.010 (61.6th percentile), scored .

Low probability of exploitation in the next 30 days, but still tracked. 62nd percentile — 61.6% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.

EPSS trend (30 days)EPSS over the last 30 days for CVE-2015-3952: held from 0.010 to 0.010.

EPSS over last 30 days · oldest: 0.010 · newest: 0.010 · change: 0.000

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2015-3952?
CVE-2015-3952 is a high-severity vulnerability in Hospira Plum A+3 Infusion System, classified under Cleartext Storage of Sensitive Information. CVSS score: 7.5/10. Published 2019-03-25.
How severe is CVE-2015-3952?
High severity. CVSS v3 base score is 7.5 out of 10.