Information disclosure in Hospira Plum A+3 Infusion System

CVE-2015-3952

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

EPSS: 0.010 (59.5th percentile) — read the EPSS interpretation.

CVSS v3 metric

CVSS v3 base score 7.5 (High). Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.

Affected products

Weakness classification (CWE)

References

  • ics-cert@hq.dhs.gov (US Government Resource, Third Party Advisory, x_refsource_MISC, Mitigation)

Frequently asked questions

What is CVE-2015-3952?
CVE-2015-3952 is a high-severity vulnerability in Hospira Plum A+3 Infusion System, classified under Cleartext Storage of Sensitive Information. CVSS score: 7.5/10. Published 2019-03-25.
How severe is CVE-2015-3952?
High severity. CVSS v3 base score is 7.5 out of 10.