Information disclosure in Hospira Plum A+3 Infusion System
CVE-2015-3952
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published · last modified .
CVSS v3 metric
CVSS v3 base score 7.5 (High). Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.
EPSS exploit prediction
EPSS 0.010, 63.6th percentile (scored ); unchanged over 30 days. About EPSS.
Affected products
- Hospira Plum A+3 Infusion System — versions <= 13.6
- Hospira Plum A+ Infusion System — versions <= 13.4
- Hospira Symbiq Infusion System — versions <= 3.13
- Pifzer Plum A\+3 Infusion System Firmware
- Pifzer Plum A\+ Infusion System Firmware
- Pifzer Symbiq Infusion System Firmware
Weakness classification (CWE)
References
- ics-cert.us-cert.gov/advisories/ICSA-15-161-01 (US Government Resource, Third Party Advisory, Mitigation)