Information disclosure in Hospira Lifecare Pca Infusion System

CVE-2015-1012

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Published · last modified .

CVSS v3 metric

CVSS v3 base score 7.5 (High). Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.

EPSS exploit prediction

EPSS: 0.008 (53.9th percentile), scored .

Very low probability of exploitation in the next 30 days; routine patching cadence is appropriate. 54th percentile — 53.9% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.

EPSS trend (30 days)EPSS over the last 30 days for CVE-2015-1012: held from 0.008 to 0.008.

EPSS over last 30 days · oldest: 0.008 · newest: 0.008 · change: 0.000

Affected products

Weakness classification (CWE)

References

Frequently asked questions

What is CVE-2015-1012?
CVE-2015-1012 is a high-severity vulnerability in Hospira Lifecare Pca Infusion System, classified under Cleartext Storage of Sensitive Information. CVSS score: 7.5/10. Published 2019-03-25.
How severe is CVE-2015-1012?
High severity. CVSS v3 base score is 7.5 out of 10.