Vulnerability in Hospira Lifecare Pca3
CVE-2014-5406
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Published · last modified .
CVSS v2 metric
CVSS v2 base score 7.6 (High). Vector: AV:N/AC:H/Au:N/C:C/I:C/A:C. This CVE predates CVSS v3 and has no v3 score from NVD.
EPSS exploit prediction
EPSS: 0.012 (68.0th percentile), scored .
Low probability of exploitation in the next 30 days, but still tracked. 68th percentile — 68.0% of CVEs in the catalogue have a lower EPSS than this one. How to read EPSS.
EPSS over last 30 days · oldest: 0.012 · newest: 0.012 · change: 0.000
Affected products
Weakness classification (CWE)
References
- www.cisa.gov/news-events/ics-advisories/icsa-15-125-01
- www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm (US Government Resource, Third Party Advisory)
- xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
- github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.js…
- af854a3a-2127-422b-91ae-364da2661108 (Third Party Advisory, US Government Resource)